
New Technologies
for Plant Science
Q. When I use software to proceed in “SetZero” procedure under “Option” manual, I got all the points piled up together at one point. What do I do?
Q. I try to do a SetZero procedure, but the system won’t respond while I am increasing the pressure. Is there something wrong?
Q. How do I eliminate air bubbles inside the Pressure Transducers?
Q. My HPFM software will not recognize the default calibration file location. I have to manually find that location every time using it.
Q. The larger 1.5mm Teflon clear Tubing does not fit to the bigger Compression Coupling Head. This is because the outer diameter of the tube is 3.2mm while the
inlet of the Coupling Head has less than 2mm.
Q. We cannot save the regression values after performing the SetZero option. A "file access denied” box appears when we try to save the regression values.
Q. Why do whole ranges disappear and change to 0.0000E +00? I thought the zero set on the transducers merely adjusted the Pressure Transducer Calibration
Factors, which appear at the top of the cal file not the actual Range Values.
Q. Is there some bug in the software causing the elimination of these ranges A0, A1 and A2, values in the cal file
Q. If there are plants whose flow rate falls between 2 adjacent ranges (ex. Red and Yellow), how should I go ahead and pick the correct range during
transienttest?
Q. How do I compare my transient test flow rate with calibration data?
Q. I try to do a SetZero procedure, but the system won’t respond while I am increasing the pressure. Is there something wrong?
Then, you need to check your PC readings on DP. In order to do so, you need to
make sure the tank is at least half full, open all the valves on manifolds except
outlet valve and open water valve as well (this usually being ignored and could
costs you hours trying to figure out why it didn’t work). Then, switch on your
HPFM system and run HPFM software. Go to “File” – “Calibrate HPFM” – “A/D
Setup” and click “Start/Stop”. Increase tank pressure to about 400 kPa and you
will be able to see a reading in the window. The DP reading should be somewhere
around 0 and PT2 reading should be around 30 mV. Then, close the inlet valve
and open outlet valve. The reading should be DP=0 mV and PT2=0.2 mV. If, in
both scenarios, the DP reading is + or – 24 out of range, you will need to re-Zero
the board. But sometimes, this is caused by just adding some new water inside the
tank. Then, you will have to bleed the tank for at least 6 hours before operating.
At last, check each pressure transducer to see if they are working properly, which
mean if they output proper readings to your voltmeter.
If you have gone through all these steps and still find no luck on running SetZero
procedure, please contact Dynamax Technical Support on HPFM for further
instruction. Please DO NOT disassemble any part of the system.
Q. How do I eliminate air bubbles inside the Pressure Transducers?
Q. My HPFM software will not recognize the default calibration file location. I have to manually find that location every time using it.
Then, double click on that file to open it in Notepad and you will see the following.
Q. The larger 1.5mm Teflon clear Tubing does not fit to the bigger Compression Coupling Head. This is because the outer diameter of the tube is 3.2mm while the inlet of the Coupling Head has less than 2mm.
Q. We cannot save the regression values after performing the SetZero option. A "file access denied” box appears when we try to save the regression values.
Q. Why do whole ranges disappear and change to 0.0000E +00? I thought the zero set on the transducers merely adjusted the Pressure Transducer Calibration
Factors, which appear at the top of the cal file not the actual Range Values.
Q. Is there some bug in the software causing the elimination of these ranges A0, A1 and A2, values in the cal file
A. This is a Common problem. First, let's check your Calibration data. Please
open your HPFM software by double clicking on the icon then go to "File" -
"View/Set Calib" to show a screen of your calibration data. Refer to Picture 1 in
the attached document. Most of your calibration data is different from this sample
so please don't intend to change yours unless instructed. Now, let's take a look at
your PT2 values: A0=-1.5219E-03 and A1=6.8965E-03. If your value is different,
manually change to the above value by double clicking on them in your
"View/Set Calib" screen and type in the correct values. Once it is changed, click
on "Save Values" then "Done". If they are the same, just click on "Done" to exit
to the main screen.
A. Use the 3cc syringe and 7/8” needle we provided in the accessory kit. Fill it
with Algaecide contained water. Eliminate all air bubbles inside the syringe and
fill the water into Transducers. Keep tapping the adapter on the transducer to
move the air bubbles to the opening of the Transducer then add water to push the
air bubble out. Make sure you DO NOT use the big 1 3/8” stainless steel needle
on this procedure or it will damage your Pressure Transducer.
A. Once you have followed the manual and copied the calibration folder onto
your hard drive, you will need to go to that folder and find the file named
“HPFM.dat”.


A. Please provide the part number of the coupling you are referring to. If it is any
of the HPCC10-50 couplings, you should be able to insert the Teflon tubing into the coupling without a problem. However, you cannot and should not force the
tubing all the way into the inside of coupling. Tubing should go into any couplings for only 2-3mm and no more. The color caps and O'rings are holding
the tubing and coupling together.
A. This is caused, I assume, by saving your regression value on your calibration
CD. If that is the case, please refer to HCFM manual Section 5.1, where it asks
you to copy your HPFM.CAL and HPFM_CAL_DATA.TXT files from the CD to your Harddrive where you install the HPFM program. If this is not the case, you
will need to check if you have specified the correct path of your HPFM.CAL file. If not, use option File-NewCalfile to specify the location of your HPFM.CAL file.
Then perform Setzero and regression. If you keep experiencing the same problem, please contact Dynamax for further assistance.
A. The set zero corrects for differences in P1 and P2 pressure in mV (not MPa as
cited below). 1 mV = 1 psi = 1/145 MPa. The HPFM software computes Flow = A0 + A1*DP + A2*DP^2. When you do a flow calibration there might be a nonzero
flow when the HPFM thinks DP =0. So A0 takes care of that. But after you do a set zero DP, the HPFM software gets a more up-to-date zero value and since
physically Flow must equal zero when DP = 0 it is necessary to set A0 = 0 for the linear ranges (usually green, orange, red and yellow). The software does not make
the adjustments for the non-linear ranges.
A. I have not seen such a bug. I have sometimes manually adjusted A0, A1 & A2 values and it seems to stick. If there are multiple cal files, you can load one that
will be used for the current session, but if you want to use the file for ALL sessions (when HPFM.exe is launched again) you have to save the cal file to make it the default file.