Specifications
Measurement type
CO2 and H2O differential, non dispersive infrared
gas analysis. Atmospheric pressure and temperature compensation.
CO2 corrected for H2O, broadening and dilution. Auto zero. Open and
Closed mode. Two experimental channels
Range measurement
CO2: 0 - 2000 ppm, H2O: 0 - 75 millibar
Resolution
CO2 0.1 ppm at 3350 ppm, H2O 0.1 millibar
Response time
Selectable between 0.1 and 10 seconds (in increments of 0.1 seconds)
Cell volume
5 cm3
Repeatability
CO2 0.2 ppm at 350 ppm
Signal Noise
CO2 typically 0.2 ppm peak to peak at 350 ppm with 1 second response time. H2O typically 0.02 millibar peak to peak at 20
millibar with 1 second response time
Differential zero drift
CO2 at 350 ppm <1ppm/hr at 25º C typically <2
ppm in 24 hours. 0.12 ppm per ºC H2O at 20 millibar typically < 0.5 millibars in 24 hours. 0.02 millibar per ºC
Span drift
CO2 <1 ppm in 24 hours at 25º C and 350 ppm. H2O <1% of f.s.d. in 24 hours at 25º C
Internal flow
control 20-600 ml min-1
Flow measurement accuracy
±1% of f.s.d.
Maximum flow rate
10 liters min-1
Pressure measurement
750 - 1250 millibar
Display
1/4 VGA color LCD
Inputs
8 scalable channels: 4 fast channels 10Hz, 4 channels 1Hz 0-5V
Data storage
Dual PC card drive (PCMCIA). Each drive compatible with a range of PC cards up to 4M Byte
Internal software
User friendly ‘graphical user interface’ familiar to PC and MAC users. User expandable and upgradable
I/O 2 RS232C serial ports for connection of mouse, printer, and PC. (One port RS-232/RS485)
Outputs
4 analogue outputs (0 - 4V) - Parallel printer port
Gas connections
Suitable for 4 mm bore tubing
Keyboard
Fully QWERTY, (wipe clean) keyboard
Operating temperature
0 to 45º C
Power requirement
230/115V 50/60Hz, 12V DC with adaptor
Dimensions
390 x 360 x 160 mm
Weight
24 lbs. (11 Kg)
|